KUV2.4 参考仪表
现场精确再校准
KUV2.4 基准仪是一种多功能设备,用于对数字系统传感器进行精确监测和现场编程,可在±30%的范围内进行重新校准。其设计符合 DVGW 和 ÖNORM 标准,是饮用水、废水和压载水系统的理想之选。审核员、检验员和实验室可以快速评估传感器是否需要重新校准或更换,确保测量准确并符合法规要求。该设备采用坚固耐用的 IP65 等级金属外壳,由可充电电池供电,适合苛刻的现场条件。其自动测量范围切换功能可确保获得可靠的结果,无需人工干预。KUV2.4 集精确性、灵活性和耐用性于一身,深受公共卫生机构和检测实验室的信赖,是现代水系统监控和传感器管理的必备工具。
Technical Information
| Configuration | Options |
|---|---|
| Housing | Aluminum, color black |
| Dimensions | 110/85 x 230 x 35 [mm] |
| Weight | 790g incl. Sensor and battery |
| Supply |
operation time 8-10 h (fully charged) plug-in power supply 100-240 V AC / 50-60 Hz with EU adapter (US, UK adapter on request) |
| Display | LCD, 2 x 16 characters with backlight function |
| Measuring range |
0.1-2000 W/m², measuring ranges with end value of 2, 20, 200, 2000 W/m2 automatic switching “overload" is displayed when the measuring range is exceeded |
| Memory | storage of the last measured value and the maximum value |
| Reference sensors |
reference sensors are equipped with 1.3 m cable and M12 socket connection is made via M12 plug on the upper right only registered sensors in the device are detected (maximum 8 possible) serial numbers of the registered sensors can be found on the sticker on the back |
| Plant sensors |
plant sensors are connected via the M12 socket on the upper left, digital or digital / analogue sensors (options Z4, Z4Y1, Z4Y2) can be programmed and calibrated necessary connection cable is included in the delivery |
| Analogue output |
voltage output: 0.1 to 4.1 V for each measuring range (0.1V=0W/m², 4.1V=2/20/200/2000W/m²) Ri = 10 kΩ, jack socket 2.5 mm in the connection compartment at the bottom of the device |
| Serial interface |
RS232 output: 9600 baud, no parity, 1 stop bit, 8 data bits, GND/TxD/RxD connection via jack socket 3.5 mm in the connection compartment at the bottom of the device |
| Degree of protection | IP 65 with sensor connected and battery / connection compartment closed |
| Temperature range | ambient temperature 0 to 30° C |
| Recalibration | latest after 12 months, according to DVGW W294 rule resp. after a useful life of 100 hours |
| Accuracy |
uncertainty to working standard: ≤ ± 1% for low pressure, ≤ ± 3% for medium pressure lamps uncertainty to internal reference: ≤ ± 1% uncertainty to primary standard DVGW/ÖNORM: ≤ ± 6.5% total uncertainty: ≤ ± 8.5% for low pressure, ≤ ± 10.5% for medium pressure lamps calibration protocol included in scope of delivery |